CX402-T205
•CX402-T215
•CX402-T230
•CX402-T405
•CX402-T415
•CX402-T430
InTemp Bluetooth Temperature Data Logger with Glycol Bottle
Important Information
For automatic downloads and remote alerts, InTemp loggers require the optional InTemp CX Gateway. MORE INFORMATION
Overview
InTemp CX402-Txxx (with Glycol Bottle) data loggers monitor temperatures in refrigerators, freezers and other temperature-controlled environments to comply with global GDP requirements. The loggers communicate wirelessly via Bluetooth to mobile devices. Using the InTemp app, you can easily view data, check logger status, set alarms, and create and share secure PDF reports for streamlined reporting and regulatory compliance. All loggers feature a built-in LCD display to view current and daily min and max temperatures, advanced audible and visual alarm capabilities for notification of temperature excursions, an internal sensor for ambient temperature monitoring, and a single-point NIST Certification of Calibration for both the logger body and probe.
With the optional CX Gateway, you can get remote notifications of temperature excursions and automatically download data from loggers within range of the device. Learn more about the CX Gateway here.
Models are available with a 2- or 4-meter external temperature probe attached to a Glycol bottle used to insulate the probe and delay thermal response times.
Features
- Integrates with mobile and cloud platforms for a 21 CFR Part 11 compliant solution
- Easy-to-read display with current and daily min/max temperatures
- Bluetooth transmission of data and alarm states
- Logger setup and download of data in seconds – without a PC or cable
- Real-time alerts and automatic data downloads with optional CX Gateway device
- +/- 0.5C accuracy for reliable temperature readings
- Battery-powered to guard against local power loss
- Several Glycol bottle/probe-only options available
- Includes 1-year NIST calibration, per USP guidelines
Who is this for?
The InTemp CX402-Txxx data logger (with probe in glycol) is right for you if:
- You work for an organization that’s required to primarily follow GDP/GMP, FDA, USP, and/or EU/CE guidelines for monitoring ambient to frozen temperatures
- Your organization is a pharmaceutical manufacturer, third-party logistics company, Life Science/Biotech storage and transportation company, or other organization looking to monitor temperature-sensitive goods
- You need a 1-Year NIST-certified data logger (probe and ambient temperature sensor), as recommended by the previously mentioned regulatory bodies
Required to follow GDP & GMP requirements, specifically:
- Probe in glycol
- GDP/GMP compliant
- Frozen to ambient monitoring
Specifications
External Temperature Probe
-
±1.0°C from -40° to -25°C±1.8°F from -40° to -13°F±0.5°C from -25° to 80°C±0.9°F from -13° to 176°F±1.0°C from 80° to 100°C±1.8°F from 176° to 212°F
-
0.024°C at 25°C0.04°F at 77°F
-
-40° to 100°C-40° to 212°F
-
CX40x-Txx and CX402-BxM: Single point NIST calibration, probe and logger bodyCX40x-VFCxxx: Single point NIST calibration, probe onlyCX403: Single point NIST calibration
-
2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable
-
Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches) long, 3.18 mm (0.125 inches) diameter
Temperature Sensor
-
-30° to 70°C-22° to 158°F
-
±0.5°C from -15° to 70°C±0.9°F from 5° to 158°F±1.0°C from -30° to -15°C±1.8°F from -22° to 5°F
-
0.024°C at 25°C0.04°F at 77°F
Logger
-
1 mW0 dBm
-
Approximately 30.5 m (100 ft) line-of-sight
-
Bluetooth Low Energy (Bluetooth Smart)
-
-30° to 70°C, 0 to 95% RH (non-condensing)-22° to 158°F, 0 to 95% RH (non-condensing)
-
1 second to 18 hours
-
Two AAA batteries, user replaceable
-
1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life.
-
128 KB (84,650 measurements, maximum)
-
Approximately 60 seconds; may take longer the farther the device is from the logger
-
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
-
9.4 x 4.5 x 2.59 cm3.7 x 1.77 x 1.02 inches
-
90.2 g3.18 oz
-
IP54
Notices
-
Passed RTCA DO160G, part 21H