Overview

The InTemp CX702 Cryogenic data logger is a low-cost Bluetooth-enabled single-use logger that records temperature data for up to 90 days. The CX702’s built-in 1.5m probe measures temperatures as low as -200°C (-328°F) and includes a protective sheath to prevent accidental cutting of the cable.

Using the InTemp® app on your mobile device, you can easily configure the CX702 with one of multiple preset configurations, or set up a custom configuration. The logger has a built-in alarm to alert you to excursions. And logger reports can be quickly downloaded to your phone or tablet and uploaded to an InTempConnect® account, where you can analyze and export data as needed.

With the optional CX Gateway, you can get real-time notification of temperature excursions and automatically download data from loggers within range of the device. Learn more about the InTemp CX Gateway.

Features
  • Single-Point ILAC/ISO17025 NIST Certificate of Calibration, valid for one year from date of purchase
  • IATA Battery compliant
  • Record up to 14,000 data points
  • Anti-cut sheath
  • Integrates with mobile and cloud platforms for a 21 CFR Part 11 compliant solution
  • Compact and lightweight – ideal for shipping pharmaceuticals
  • No need to open containers to access data
  • Ensures secure data via encrypted technology
Specifications

Temperature Probe

Sensor Range
-200° to 50°C (-328° to 122°F); To ensure insulation integrity, avoid flexing the wire during exposure to temperatures below -100°C (-148°F)
Accuracy
±1.8°C from -200° to -90°C (±3.2°F from -328° to -130°F)
±1.0°C from -90° to -10°C (±1.8°F from -328° to 14°F)
±0.5°C from -10° to 50°C (±0.9°F from 14° to 122°F)
Resolution
0.013°C at 25°C (0.023°F at 77°F)
NIST Calibration
Single-point NIST calibration, probe only @ -195C
Cable Length
1.5 m (4.92 ft)
Probe Dimensions
Food-grade stainless steel probe with blunt tip, 31.75 mm (1.25 inches) long, 4.78 mm (0.188 inches) diameter

Logger

Logger Operating Range
-30° to 50°C (-22° to 122°F)
Radio Power
1 mW (0 dBm)
Transmission Range
Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard
Bluetooth Low Energy (Bluetooth Smart)
Type
Single use
Duration
90 days
Shelf Life
Up to 1 year
Logging Rate
1 second to 18 hours
Battery
CR2450 3V lithium, not replaceable. Battery life is 1 year, typical with logging interval of 1 minute.
Faster logging intervals, remaining connected with the InTemp app, excessive downloads, and paging may impact battery life. Battery lasts for duration specified when logger is used within the 1-year shelf life.
Memory
16,000 measurements
Full Memory Download Time
Approximately 30 seconds; may take longer the farther the device is from the logger
Dimensions
6.9 x 4.5 x 1.1 cm (2.71 x 1.76 x 0.42 inches)
Weight
25.5 g (0.90 oz)
Environmental Rating
IP65

Notices

Passed RTCA DO160G, part 21H
Image
CE Compliance Mark
Image
United States FCC Mark
Image
Australia-New Zealand Regulatory Compliance Mark (RCM)
Documents

InTemp CX600 Dry Ice/CX700 Cryogenic Logger Manual

InTemp CX600 Dry Ice and CX700 Cryogenic loggers are designed for cold shipment monitoring and have a built-in external probe that can measure temp

View View

Quick Start for the InTemp CX600 and CX700 Series Loggers

A quick start guide on how to set-up, configure, start, and deploy your CX600 or CX700 series logger.

View View

InTemp System User’s Guide

CX series loggers, the InTemp® app, and InTempConnect™ web-based software work together to form the InTemp temperature monitoring solution.

View View