Custom pH Buffers to Calibrate the HOBO MX2501 pH Logger
The HOBO MX2501 pH logger calibration procedure is only designed to work with National Bureau of Standards (NBS) pH buffers 4.01, 7.00, and 10.00.
Read tech note
Using a Three-Point pH Calibration versus a Two-Point pH Calibration
In order to guarantee the accuracy of your readings, it is necessary to bracket the expected pH values with at least two buffers.
Read tech note
Calibrating for pH at Temperatures Other than 25°C
The observed pH of the buffers (and samples) varies depending on temperature.
Read tech note
How to deploy a data logger in turbulent water
See image for a quick deployment idea for an underwater logger, such as the HOBO Water Temp Pro v2 or the HOBO U24 Conductivity logger.
Read tech note
Barometric Compensation Method
Barometric Compensation Method - Technical Details
This section presents the details of how the Pressure values are computed, and how Pressure and Temperature values
Read tech note
HOBO U20 Stainless Steel Data Logger - sensor location
See attached diagram for sensor location in HOBO U20 stainless steel data loggers.
Read tech note
Pendant Optics
Attached please find an image that shows the location of optic LEDs in Pendant (UA-series) loggers. All Pendant optics are located in the same location on all models.
Read tech note