W-CTD-01
•W-CTD-02
•W-CTD-03
•W-CTD-04
CTD Sensor
Important Information
Requires a HOBO MX801 or MX802 Data Logger, the HOBOconnect app, and a compatible mobile device or Windows computer. System requirements for the app can be found at the bottom of our HOBOconnect software page.
Overview
The CTD sensor is an interchangeable sensor that works with HOBO MX800 Series Water Loggers to measure conductivity, temperature, and water level. With this sensor attached to a HOBO MX800 Series Water Logger, the logger can also calculate specific conductance, salinity, and total dissolved solids (TDS).
The CTD sensor can be attached directly to the fully submersible MX801 model, or attached via cable to the direct read MX802 model, which doesn’t require having to pull the sensor out of the water to download data.
Features
- 4-electrode conductivity sensor provides wide measurement range for monitoring freshwater or saltwater
- Conductivity sensor is centrally located to minimize effects from stilling wells or protective guards
- Durable ceramic water level sensor, located in the sensor tip for accurate measurement of shallow water levels
- Get water level directly with the MX802, no need for post-processing
- 1” diameter for use in narrow wells (with cable to MX802 above the well)
- Temperature sensor placement near the sensor tip for fast response time
- Specially treated stainless steel holds up in saltwater
Specifications
Conductivity Measurements
-
Electrical conductivity and specific conductance: 0 to 100,000 µS/cmSalinity using PSS-78: 2 to 42 PSUTotal Dissolved Solids (TDS): 0 to 100,000 mg/L
-
Electrical conductivity: 50 to 80,000 µS/cm5° to 35°C41° to 95°F
-
Conductivity: +/-2% or 15 µS/cm whichever is greaterSalinity: +/-2% of reading or 0.1 PSU, whichever is greater
-
Conductivity:0.1 µS/cm from 0 to 1,000 µS/cm1 µS/cm from 1,000 to 10,000 µS/cm10 µS/cm from 10,000 to 100,000 µS/cmTDS: 0.1 mg/LSalinity: 0.01 PSU
-
1 second to 90% at a stable temperature
Absolute Pressure and Water Level Measurements W-CTD-01
-
0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude
-
69 to 207 kPa (10 to 30 psia), 0° to 40°C (32° to 104°F)
-
310 kPa (45 psia) or 18 m (60 ft) depth
-
Typical error: ±0.05% FS, 0.5 cm (0.015 ft) waterMaximum error: ±0.1% FS, 1.0 cm (0.03 ft) water
-
±0.3% FS, 0.62 kPa (0.09 psi) maximum error
-
<0.02 kPa (0.003 psi), 0.21 cm (0.007 ft) water
-
<1 second to 90% at a stable temperature
Absolute Pressure and Water Level Measurements W-CTD-02
-
0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) of water depth at sea level, or 0 to 33.6 m (0 to 111 ft) of water at 3,000 m (10,000 ft) of altitude
-
69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F)
-
500 kPa (72.5 psia) or 40.8 m (134 ft) depth
-
Typical error: ±0.05% FS, 1.5 cm (0.05 ft) waterMaximum error: ±0.1% FS, 3.0 cm (0.1 ft) water
-
±0.3% FS, 1.20 kPa (0.17 psi) maximum error
-
<0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) water
-
<1 second to 90% at a stable temperature
Absolute Pressure and Water Level Measurements W-CTD-03
-
0 to 850 kPa (0 to 123.3 psia); approximately 0 to 76.5 m (0 to 251 ft) of water depth at sea level, or 0 to 79.5 m (0 to 262 ft) of water at 3,000 m (10,000 ft) of altitude
-
69 to 850 kPa (10 to 123.3 psia), 0° to 40°C (32° to 104°F)
-
1,200 kPa (174 psia) or 112 m (368 ft) depth
-
Typical error: ±0.05% FS, 3.8 cm (0.125 ft) waterMaximum error: ±0.1% FS, 7.6 cm (0.25 ft) water
-
±0.3% FS, 2.55 kPa (0.37 psi) maximum error
-
<0.085 kPa (0.012 psi), 0.87 cm (0.028 ft) water
-
<1 second to 90% at a stable temperature
Absolute Pressure and Water Level Measurements W-CTD-04
-
0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) of water depth at sea level, or 0 to 7 m (0 to 23 ft) of water at 3,000 m (10,000 ft) of altitude
-
69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F)
-
310 kPa (45 psia) or 18 m (60 ft) depth
-
Typical error: ±0.075% FS, 0.3 cm (0.01 ft) waterMaximum error: : ±0.15% FS, 0.6 cm (0.02 ft) water
-
±0.3% FS, 0.43 kPa (0.063 psi) maximum error
-
<0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) water
-
<1 second to 90% at a stable temperature
Temperature Measurements (All Models)
-
-20° to 50°C (-40° to 122°F), non-freezing water
-
±0.15°C (±0.27°F) from 0° to 50°C (32° to 122°F)
-
0.01°C at 25°C (0.018°F at 77°F)
-
3 minutes to 90% in water (typical)
-
<0.1°C (0.18°F) per year
Sensor
-
2.5 cm (1 inch) diameter, 18.7 cm (7.37 inches) length
-
177g6.24 oz
-
Passivated 316 Stainless steel housing rated for use in saltwater, Viton and Buna-N O-rings, PET sensor connectorDepth sensor: ceramicConductivity sensor: PET, platinum plating on electrodesNote: Sensor should be mounted so that it is not in contact with other metals
-
IP68; Waterproof to pressure sensor burst pressure rating
* Water Level Accuracy: With accurate reference water level measurement, known water density, and a stable temperature environment. System Water Level Accuracy equals the sum of the Barometric Water Level Accuracy plus the selected CTD sensor Water Level Accuracy.
** Raw Pressure Accuracy: Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors.
*** Changes in Temperature: Allow 20 minutes in water to achieve full temperature compensation of the pressure sensor. There can be up to 0.5% of additional error due to rapid temperature changes.
Compatible items
Data Logger(s)
Antifouling Protective Guard
Accessories
Communication Cables
Calibration Solution
Accessories
What's in the box
- CTD Sensor