T-DS-STS2101S-025
pHionics STs Series® ORP Sensor
3rd party Sensor, RX station-compatible
Important Information
25-foot cable included. Manufactured and distributed by a third party, pHionics, this sensor can deliver real-time data to the LI-COR Cloud remote monitoring platform when connected to a cellular HOBO RX station via the Analog Module or with a HOBOnet wireless sensor network's Analog Mote.
Overview
The pHionics STs Series™ ORP sensor (sonde) provides reliable ORP and measurements to monitor bacterial activity, sanitization, or heavy metal pollution. This sensor may be connected to an RX station using an an Analog Mote and HOBOnet Manager (wireless) or an Analog Module (wired).
Accurate and low-maintenance operation is ensured with noise-resistant 4-20 mA analog signals and a large-volume, double-junction electrode. Typical electrode life is 2-4 years and, when needed, a replacement is easily installed thanks to pHionics’ patented pHiConn™ Connector System. In addition, the electronics are isolated to prevent ground loops and proprietary signal conditioning virtually eliminates background noise.
Durability is guaranteed with chemical-resistant 316 Stainless Steel (titanium optional) and high-quality double O-ring seals that last many years, even in harsh conditions. When combined with the compact design (0.75 in. or 19 mm diameter), this ORP sensor proves to be compatible with a wide range of applications including aquaculture, environmental monitoring, and pollution control.
If you need a sensor that works anytime, anywhere, for many years, then the pHionics STs Series ORP sensor is here for you.
Features
- Combination double-junction electrode with large, gel-filled reference
- Secondary 0-50°C temperature sensor
- 4-20 mA signal capable of transmission up to 3 miles (with proper wire gauge)
- Built-in electrical isolation to prevent ground loops
- Chemical-resistant PVDF and 316 Stainless Steel housing (Titanium optional)
- Narrow diameter (0.75 inches or 1.9 cm)
- Removable electrode guard to reduce debris or algae buildup and protect from damage
- Fully submersible up to 70 psi
- Shielded, Kevlar®-reinforced, water-blocked polyurethane cable
- Reversible Input Protection and Operation
- 2-second warm-up time and low power usage to avoid battery drain of datalogger or RTU
- Signal conditioning to remove background noise
Specifications
ORP Sensor
| Output | 4-20 mA |
| ORP (redox) Range | -1000 to +1000 mV |
| Operating Temperature | 0-70°C |
| Pressure Range | 0-70 psi |
| Electrode Type | Platinum sensing element Silver/silver chloride gel-filled reference |
| Transmitter Type | 2-wire |
| Automatic Temperature Compensation | None |
| Isolation | 600 VDC |
| Power Supply Voltage | 8-40 VDC |
| Warm-up Time | 2 seconds |
| Loop Impedance (Max) | 200 Ohms at 12 VDC 800 Ohms at 24 VDC 1600 Ohms at 40 VDC |
| Response time | 95% < 20 seconds |
| Wetted Materials | 316 Stainless Steel, PVDF, Viton™ |
| Length | 14.25 in (36.20 cm) |
| Width | 0.750 in (1.9 cm) maximum |
| Weight (No cable) | 0.50 lbs (0.227 kg) |
| Linearity | ± 4 mV |
| Accuracy | ± 4 mV* |
| Sensitivity | ± 1 mV |
| Stability | ± 2 mV |
| Repeatability | ± 10 mV |
*After proper calibration
Temperature Sensor
| Output | 4-20 mA |
| Temperature Range | 0-50°C |
| Accuracy | +/- 1°C |
| Linearity | +/- 0.5°C |
| Power Supply Voltage | 7-40 VDC |
Cable
| Cable Material | Polyurethane |
| Type | 4 conductor, Kevlar® reinforced, water blocked, shielded, twisted-pair wiring |
| Maximum Distance | 3 miles |
| Resistance per 1000 ft (304.8 m) | 26 Ohms |
| Replaceable? | Yes |