StowAway IS Temperature Data Logger

Part # TBIB-40+75-IS
 
This product is discontinued.
 

Overview

The StowAway-IS Temperature data logger is rated intrinsically safe for recording temperatures in hazardous locations. Its small size, wide temperature range (-40 to 75°C), and splash-proof design make it ideal for deployment in harsh industrial environments. Intrinsically safe rating (by US standards): IS, Class I, II, Division 1, Groups A-G, Temperature Code T4 (<=135°C); NI, Class I, Division 2, Groups A-D, Temperature Code T4 (<=135°C); S, Class II, Division 2, Groups F and G; as specified in the US National Electric Code and approved by Factory Mutual.

Highlighted Features

  • Certified intrinsically safe for monitoring hazardous locations where combustible gas and/or dust may be present

In what environment does this data logger operate?

This data logger operates in indoor and outdoor environments.

What measurements does this data logger support?

The TBIB-40+75-IS data logger supports the following measurement: Temperature

The country of origin for this product is the United States. To see the full specifications for this product, please see our product manual found under the documentation tab if available.

Operating Temp Range*: -40°C to 75°C (-40°F to 167°F)
Operating RH Range:
at 0-100% RH and <30°C (86°F): continuous exposure
at 0-95% RH and >30°C (86°F): continuous exposure
at 95%-100% RH and >30°C (86°F): max of 10 weeks

Note: exposure to continuous moisture above 30°C (86°F) for more than 10 weeks may affect temperature accuracy

Time Accuracy: ~100 ppm (±1 minute per week) at 20°C (68°F), see Plot A for full
temperature/time dependence

Temperature Accuracy: ±0.46°C (±0.72°F) at 20°C (68°F)

Temperature Resolution: 0.35°C (0.63°F) at 20°C (68°F)

Response Time: <60 minutes in still air

Recommended Storage Temperature: -40°C to 50°C (-40°F to 122°F); storage outside this range will reduce battery life

Measurement Capacity: 32K 8-bit measurements

Data Offload Time: 8K in 90 seconds, 32K in 6 minutes

Size: 1.5" diameter, 0.8" height, mounting bail hole diameter 3/16" (5mm)

Weight: 1.1oz (32g)

Battery: 1/10 D, 3.6 Volt Lithium, non-replaceable

Battery Life: 5 years continuous use with a logging interval ≥10 minutes and multiple sampling disabled

Environmental Rating: Tested to NEMA 6 (IP67); not intended for deployment in water or extended exposure to epoxy-permeating vapors

Intrinsically Safe Rating: IS, Class I, II, Division 1, Groups A-G,
Temperature Code T4 (£135°C); NI, Class I, Division 2, Groups A-D,
Temperature Code T4 (£135°C); S, Class II, Division 2, Groups F and G

*Measurement range may differ from operating range due to
precision calibration process by up to ±2.0°F (1.1°C)

The following are associated resources for this product. In general users of this product can gain insight into its possible use by utilizing the resources below.