If you are using both TRMSA-D FlexSmart modules and smart sensors with your HOBO RX3000 station, you are limited to a maximum of three TRMSA-D modules and any combination of smart sensors to reach a total of 13 data channels.

For example:

  • If there are three TRMSA-D modules in use with the RX3000 station (each module has two channels for a total of six in all), then this will leave seven additional data channels for smart sensors

Similarly:

  • If there are two TRMSA-D modules in use with the RX3000 station, this will leave nine additional data channels for smart sensors.

Refer to the smart sensor manuals at www.onsetcomp.com/support/manuals for the number of data channels used by a specific smart sensor. If you are not using smart sensors with the HOBO RX3000 station, then you are limited to a maximum of four TRMSA-D modules. Exceeding these guidelines may result in erroneous readings.

Categories

Tech Note
Getting Started