CX Gateway System Sensor Alarm Information

The InTempConnect notification system is a great way to be alerted if your refrigerator or freezer isn't keeping the right temperature. Once a sensor alarm is breached, the logger connects to the CX Gateway device and the notification is sent. 

It's important to note, for six hours after that, the logger cannot reconnect to the Gateway. If a sensor alarm is breached within that six-hour period, the logger will not connect and no notification will be sent. Even after the six-hour window is complete, alarms that might have occurred within that time will not result in a notification. Any new alarms that occur after the six-hour window will be processed normally.    

Measurement and accuracy