Part Number - TBI32-05+37
HOBO StowAway TidbiT Temperature Data Logger
-5° to 37°C Temperature (1,000 ft.)
Discontinued
Important Information
HOBO data loggers require HOBOware software (USB interface cable included w/CD purchase) and a communications device. See compatible items below.
Compatible with
Overview
This tiny completely sealed underwater temperature data logger utilizes optic communications for launch and readout. This model records 32K of temperature measurements from -5°C to 37°C (24 °F to 99°F). Waterproof to 333m (1000ft), StowAway TidbiT units should not be used in wet (>90% RH) environments and water temperatures higher than +30°C for more than eight weeks cumulatively.
Features
- Waterproof to 1000 feet
- Completely sealed in epoxy; very durable
- User-selectable sampling interval: 0.5 seconds to 9 hours, recording times up to several years
- Blinking LED light shows if temperature goes out of user-determined limits
- Precision components eliminate the need for user calibration
- Programmable start time/date
- Triggered start with coupler or magnet
- Memory modes stop when full or wrap-around when full
- Nonvolatile EEPROM memory retains data even if battery fails
- Multiple sampling with minimum, maximum or averaging
- Blinking LED light confirms operation
- Time accuracy: ±1 minute per week at +68°F (+20°C)
- Mounting tab
- Compliance certificate available
- NIST-traceable temperature accuracy certification available
- 5 year non-replaceable battery (typical use :16 three-month deployments in water (+35°F to +80°F) with 4 minute or longer intervals (no multiple sampling), 1 offload per deployment.)
Specifications
full dependance shown in Plot A
Depth: Rated to 1000 feet
Range†: -04° to 37°C (24° to 99°F)
Accuracy: ±0.2° at 20°C (±0.4° at 70°F)
Resolution: 0.16° at 20°C (0.29° at 70°F)
Capacity: 32,520 measurements
Operating temperature range: -20° to 50°C
Battery line typical use: 5 years
Size: 1.2 x 1.6 x 0.45"
Weight: 0.5 oz
†Specified range is narrower than nominal range due to precision calibration process. Using TidbiT Temp loggers in wet environments (>90% RH) over 86°F (30°C) for extended periods of more than 8 weeks cumulative may lead to premature failure. For applications over 30°C use the HOBO® Water Temp Pro v2.