U-DTW-1 Waterproof Shuttle LED Error Codes
U-DTW-1 Waterproof Shuttle LED Error Codes
|
Most Frequent Errors (*) |
LED Error Code |
Description |
Likely Cause of Problem |
|
11 |
Logger communication failure during offload |
Logger moved during offload |
|
|
* |
12 |
No logger found |
Clean optic surfaces, check logger battery |
|
* |
13 |
Shuttle is full |
Offload and relaunch shuttle |
|
* |
14 |
No tags found in logger or invalid logger memory size |
Logger was not launched with HOBOware 2.2+ Unlikely Cause: Logger has more than 64K of memory and is not supported by this version of Shuttle Firmware |
|
21 |
Logger communication failure during launch (data offloaded ok) |
Logger moved during offload |
|
|
22 |
Invalid re-launch parameters/first 512 bytes of header could not be offloaded |
Possibly a corrupted header in the logger |
|
|
23 |
Logger header can't be verified, can't launch |
Logger moved during offload |
|
|
24 |
SHUTTLE_INFO tag version incompatibility, can't launch |
A unsupported future version of HOBOware was used to launch the logger in question, look for a Shuttle firmware update |
|
|
25 |
RELAUNCH_COMMANDS tag version incompatibility, can't launch |
A unsupported future version of HOBOware was used to launch the logger in question, look for a Shuttle firmware update |
|
|
* |
33 |
Low shuttle battery voltage |
Shuttle batteries are below 2.2 volts, replace and relaunch |
|
* |
34 |
Invalid shuttle clock time |
Shuttle was not launched after power reset |
|
42 |
Logger failed communications test |
Check logger alignment, clean optical surfaces |
|
|
44 |
Atmel memory test failed |
Bad memory, return to Onset for memory replacement |
|
|
Yellow LED is on continuously |
Switch is stuck on |
Magnet is in coupler is too powerful/reed switch is too sensitive |
|
|
All three LEDs Blink simultaneously |
Power reset |
Loose connection on PCB or something is shorting on the PCB |
|
|
Green and Yellow LEDs alternately blink |
Shuttle is in Firmware Update Mode and has enumerated with a USB port on a PC |
Logger was powered up with the reed switch activated. Remove coupler and cycle power. |
|
|
Green and Yellow LEDs blink simultaneously |
Shuttle is in Firmware Update Mode and but is not connected to a USB port |
Logger was powered up with the reed switch activated. Remove coupler and cycle power (batteries and USB cable). |