A single-channel logger with internal temperature sensor, the H02-001-02 records up to 7943 measurements.
Features:
8-bit resolution
Programmable start date/time
Programmable sampling rate (0.5 second to 9 hours)
Environment:
The H08-001-02 Data Logger is for use in Indoor environments
Measurements:
The H08-001-02 Data Logger supports the following measurements : Temperature
CE Compliance
Onset products have been CE compliant for many years. Recently the European Union changed the CE regulations, effective July 20, forcing Onset to either change the circuitry in some products, or choose to no longer list them as CE compliant. In most cases we offer CE compliant alternative products.
Rest assured Onset has placed a high priority on making all current products CE compliant before year-end. If you are not shipping this item to a EU country after July 20, please disregard this message.
This product is not currently CE compliant. If you are shipping to a EU country, Onset recommends using the CE compliant parts: U10-001 or U12-001
Operating range (logger): -20°C to +70°C (-4°F to +158°F), 0 - 95% RH non-condensing
Time accuracy: approx. ±1 minute per week (±100 ppm at +20°C or +68°F), full dependance shown in Plot A
Measurement capacity: 7944 measurements total, stored in non-volatile memory
Size: 2.4" x 1.9" x 0.8"
Weight: approximately 1 oz.
Battery: CR-2032 (lithium) user-replaceable
Battery life (continuous use): 1 year
Storage temperature: -40°C to +75°C (-40°F to +167°F)
All HOBO and StowAway data loggers shut off when the memory is full (unless wrap around is selected), or the data is downloaded to a computer. Once the data is downloaded to the computer, the data logger is in off mode and will need to be relaunched to continue data collection. If the data logger you are using is shuttle-compatible and you are downloading to a shuttle, the shuttle will automatically relaunch the data logger using the previous settings.
A discharge of static electricity may cause a running data logger to stop recording. Precautions should be taken to prevent this while handling the data logger. Discharges may occur in the following situations:
When there is a charge difference between the data logger and a cable being connected to it or a person touching an exposed jack.
HOBO/StowAway data loggers shut off when the memory is full (unless wrap around is selected), or the data is downloaded to a computer. Once the data is downloaded to the computer, the data logger is in off mode and will need to be relaunched to continue data collection. If the data logger you are using is shuttle-compatible and you are downloading to a shuttle, the shuttle will automatically relaunch the data logger using the previous settings.
Relaunching a logger before the data has been saved
Technical problem with the logger itself.
The following section discusses data loss and its causes; it also discusses methods of retrieving lost data, and how to insure against data-loss problems.
If possible, remove the board from the case and the battery from the board. Dry the board completely using a hair dryer. If the board was exposed to salt water, rinse the board with fresh water and then dry the board. You may try inserting a new battery and downloading the data. If you are unable to access the data, or see any sign of corrosion or tarnishing on the board, contact the place of purchase for a return number. The data may be retrievable, and the data logger may be repairable. Charges will apply and are dependent on the amount of damage.
As a general rule, no. This is dependent on the data logger you are using and the chemical and/or concentration of the chemical that the data logger is exposed to. Many chemicals will have an adverse effect on the data loggers and/or the sensors that the loggers utilize. Contact your authorized sales representative with questions regarding specific chemical contact with the data logger that you are using.
If a data logger is disconnected immediately after selecting launch (or start), the information that is stored on the data logger's header may become corrupted. To correct the problem, simply relaunch the data logger. Allow the launch parameters to be sent to the data logger before disconnecting the data logger from the interface cable. Allow the data logger to record a few measurements. Download the information from the data logger to see if the DDDDDDs in the Description field were overwritten.
Onset offers a USB Serial Adapter which provides a simple way to connect HOBO and StowAway® data loggers to available USB ports on your PC.
Onset also offers a USB solution with the U12 Family of data loggers that interface directly with the USB port on your computer.
Chances are it will not work. In all the USB-to-Serial Adapters we have tested, we have only found one that works reliably: the Keyspan serial adapter listed below.
There is a small bug in early versions of the HOBO Water Temperature Pro data loggers that result in premature starting in some cases (when using the delayed-start feature). With the affected data loggers there is a 0.4% chance they will start up to 4.3 minutes early, and a 0.4% chance they will start up to 18.2 hours early. If premature delayed-start occurs, all measurements are taken at the correct sampling rate and stored with the actual times they were taken, but there will be extra readings at the beginning of the data file.
Open HOBOWare, plug the logger back in, then click on Start->Settings->Control Panel->System. Click on Hardware->Device Manager and look for the logger listed under: Other Devices or Human Interface Devices — there could be a minus sign beside one of the categories or a yellow exclamation point through the device. The description listed should indicate if it is a HOBO or not. If it is not a HOBO, do not uninstall it, as it may be another USB device. Delete the "other device," then unplug the logger, wait 15 seconds and plug it back in.
Check both the logging and the sampling interval to make sure they are greater than one-minute intervals. For example, with a sampling interval of one second, the HOBO Weather Station doesn't sleep and there is a constant drain to the batteries. Make sure that you have protected the board from condensation and insects.
The Optic and Tidbit data loggers must be inserted into the Base Station in order for the computer to detect them; it will not detect the Base Station alone. Check to make sure that the coupler is positioned properly and the magnet end is closest to the data logger, not the Base Station or the Optic Shuttle. It is important to keep the Optics and Tidbits free from obstructions such as dirt and algae. Do not use any chemicals to clean the data loggers. Clean the Optics with warm water and mild dish detergent.
No. The data is stored on non-volatile EEPROM. For data loggers with external batteries: change batteries and offload data, then relaunch normally. For data loggers with internal batteries: as long as the EEPROM has not been damaged, the data can be retrieved by Onset Computer Corp. Please contact the place of purchase for a return authorization number and request data back service.
If the sensor is removed for more than 5 minutes while logging, the data logger will no longer recognize the sensor when it is reconnected. At that point it is necessary to relaunch the data logger in order to initialize the sensor, as the data logger will not recognize the sensor
The first and last event on your datafile is a stamp of the start and end time. For an accurate account of events, these two must be deducted from your totals.
The TL-5186 lithium batteries build up an internal passive resistance over a period of non-use. This build-up will interfere with the voltage supply to the data logger. To clear the resistance, remove the battery from the data logger and short the two battery prongs (by placing a small metal object over them), for about a half second. Place the battery back in the data logger. If the data logger is receiving power from the battery, the LED light will blink a few times rapidly upon contact. Attempt to launch the data logger again.
For input voltage to the data logger greater than 2.5 volts DC, you will need to make a voltage divider. A voltage divider will drop the voltage of the input signal to the level the data logger can measure. First determine the needed attenuation ratio; divide the input volts by the output volts to the data logger: A = input volts/2.5
A discharge of static electricity may cause a running data logger to stop recording. Precautions should be taken to prevent this while handling the data logger.
Discharges may occur in the following situations: When there is a charge difference between the data logger and a cable being connected to it or a person touching an exposed jack.
The H8 series data loggers measure analog voltages with an 8-bit A/D convertor, therefore any sensor's input will have not more than its range divided by 256 values. For linear inputs, like a raw voltage input, the value for each step is equal. For nonlinear inputs, like temperature, the values of the steps are not equal. For instance, with the built-in temperature sensor on H08-001 through 004, the step value at 70° is 0.7°, while at about 175°, the step value is 2°. This step value is specified in our literature using the term "resolution."
The CO data logger should be checked, at a minimum, on a yearly basis. Proper testing requires a CO testing chamber. If you do not have access to one, you may contact the place where you purchased the HOBO CO data logger and return it for testing. If the data logger is not reading within specification, the sensor will need replacement. Please note that the sensor should not be removed from the board for more than 1-2 hours. If the sensor has been removed for a period of time, allow 24 hours from the time it is replaced for the readings to stabilize.
The maximum recommended impedance for analog sources is 2.5 K. This value is calculated based on the maximum leakage current of the input pin. The leakage current is 100 nA max., and the analog input voltage cannot be varied by more than 1/4 LSb or 250
All U10 and U12 data loggers are fully compatible with the Real-Time Alarm Software. The data loggers can connect to a USB port on the computer or remotely through an Ethernet network using a Keyspan USB Server. HOBO Weather Station (H21) and HOBO Energy Logger (H22) data loggers connected to a serial port on the PC are also compatible. The remote communication products for HOBO® Weather Stations are not supported.
For best performance, connect the USB cable directly to a port on your computer. The USB ports on your computer may be more reliable than hubs or keyboard ports.
A data logger (or datalogger), commonly referred to as a 'HOBO', is an electronic instrument that records measurements (temperature, relative humidity, light intensity, on/off, open/closed, voltage, pressure and events) over time. Typically, data loggers are small, battery-powered devices that are equipped with a microprocessor, data storage and sensor. Most data loggers utilize turn-key data logger software on a personal computer to initiate the logger and view the collected data.